Portable oil contamination detection instrument 0203
Oil particle counter
Product performance:
This instrument can monitor and analyze the particle size, cleanliness, and pollutants of oil; Detection and evaluation of insoluble particulate impurities in various aircraft fuel oils, aviation lubricants, and phosphate hydraulic oils (corrosive), as well as military support equipment and its daily maintenance and upkeep; Wear testing in combat readiness equipment components; Test for insoluble particles in pure solutions and ultrapure water.
This instrument adopts the UK puluody Oil Particle Detection Technology - "Light Resistance Measurement Particle", and can be equipped with multiple standards according to user requirements.
Technical parameters:
·Customized requirements: testing requirements for various lubricants;
·Laser sensing detector: 7th generation dual laser narrow light detector;
·Testing software: P6.4 analysis testing software integrated version&PC version;
·Control method: Integrated industrial computer control or industrial PC control;
·Evaluation criteria: Meet standards such as ISO4406-87/99, NAS1638, SAE AS 4059F, GJB420B, etc;
·Testing standards: DL/T432, GJB380, ISO11500
·Calibration standards: JJG1061, MTD/ISO11171, ACFTD/ISO4402, GB/T18854
·Latex ball method or ASTM F658-87 (PSL latex ball)
·Operation mode: Color LCD touch screen operation;
·Special detection: Custom detection 1-100 μ M particles, 0.1 μ M arbitrary detection;
·Sensitivity: 1 μ M (ISO 4402) or 4 μ M (C) (ISO 11171, GB/T18854);
·Number of channels: 1000, can count particles in any 16 particle size ranges;
·Coincidence accuracy: 10000 particles/mL (5% coincidence error);
·Resolution: 90% (60ml/min)
·Accuracy:<3%
·Injection volume: 1-200ml, can be set arbitrarily;
·Volume accuracy: better than ± 1%;
·Online pressure: 0-0.6MPa (excluding pressure reducing valve); 0-31.5MPa (including pressure reducing valve);
·Test viscosity: offline 0-99mm2/s; Online 0-500mm2/s;
·Sampling flow rate: 5mL/min~60mL/min;
·Cleaning flow rate: 5mL/min~150mL/min;
·Test time: can be set arbitrarily;
·Sample processing: The sample to be tested does not require pre-treatment;
·Evaluation level: Oil contamination level can be directly evaluated based on NAS1638, ISO 4406, etc;
·Data output: Can display particle size, quantity, curve, and distribution