The differentiation and integration in the oil particle counter detection report correspond to different standards. Specifically, as follows:
1. Differential standard: Differential standard refers to the classification of particle pollution levels based on the NAS1638 standard. The NAS1638 standard divides the maximum allowable particle count in oil within the size range of 5-15, 15-25, 25-50, 50-100, and>100um into 14 pollution levels. The measured particles in each size range often have different levels, and generally one level is taken as the oil pollution level.
2. Integral standard: The integral standard refers to the classification of particle concentration levels according to the ISO4406 standard. The ISO4406 standard is divided into 26 levels based on the size of particle concentration. The country has made modifications and improvements to the ISO4406 standard. The revised ISO4406:1999 stipulates that for pollution levels counted using automatic particle counters, three particle concentration levels of>4um,>6um, and>14um are used to represent.
In summary, the differentiation and integration in the oil particle counter detection report correspond to different aspects of the NAS1638 and ISO4406 standards, respectively. The differentiation corresponds to the pollution level, while the integration corresponds to the particle concentration level.
The detection results of PLD-0201, PMT-2, and PLD-0203 oil particle counters can be output according to standards such as NAS 1638, ISO 4406, SAE 4059E, SAE 4059F, GJB 420A, and GJB 420B.
The puluody oil particle counter adopts puluody technology - "light resistance measurement of particles", and adopts the classic methods of NAS1638 and ISO4406 in the oil industry. It can also be equipped with multiple standards according to user requirements.
By using precision plunger pumps and ultra precision flow electromagnetic control systems, dual control is achieved for constant injection speed and precise injection volume. The sampling volume is set arbitrarily from 1ml to infinity, ensuring accuracy and accuracy.
Built in statistics, particle size curve, and pulse resistance value, the channel particle size value can be set.
Integrated fully automatic sampling device, equipped with pressure system and stirring device, enables the instrument to achieve sample degassing, uniform and high viscosity sample detection.
Adopting a large LCD display, touch screen menu operation, keyboard and touch dual input, with a beautiful appearance and complete functions.
The data processing function is powerful and rich, and it can draw distribution histograms based on the oil level and data provided by the user.
Built in operating system and micro printer, allowing for direct testing and printing without the need for external computers and printers.
Equipped with a standard serial RS232 port, it can be connected to an external computer to store detection results, facilitating data classification and retrieval.
It can be calibrated according to standards such as ISO11171 and ISO4402.